Mon - Sat 07AM - 07PM +91-9585599560exports@freedomophthalmic.com

TORIC HYDROPHOBIC IOL

Freedom TORIC     Freedom TORYC     Freedom TORIC MULTIFOCAL     Freedom TORYC MULTIFOCAL     Freedom TORIC EDOF     Freedom TORYC EDOF     Freedom TRITORIC     Freedom TRITORYC

Treat Astigmatism

Astigmatism is very common before and after cataract surgery. Almost 40% of people who are candidates for cataract surgery have at least 1.0 diopter (D) of astigmatism, which is enough to cause noticeably blurred vision without eyeglasses or visual aids.

With Astigmatism

With Freedom TORIC

The correction of ASTIGMATISM OF THE CORNEA is possible by the insertion of a TORIC IMPLANT that generates an astigmatism in opposite direction of the corneal one. It allows to neutralize the entire eye astigmatism. The toric implant compensates astigmatism induced by corneal.

Why choose the FREEDOM TORIC IOLs?

The premium design of our TORIC Hydrophobic Lenses provides distance vision for ASTIGMATIC CATARACT PATIENTS who desire spectacle independency.

Our TORIC IOL is available in 8 different models (in Clear and Yellow, Preloaded or non-preloaded)

  • FREEDOM TORIC, Monofocal Toric with Clear Hydrophobic
  • FREEDOM TORYC, Monofocal Toric with Yellow Hydrophobic
  • FREEDOM TORIC MULTIFOCAL, Multifocal Toric with Clear Hydrophobic
  • FREEDOM TORYC MULTIFOCAL, Multifocal Toric with Yellow Hydrophobic
  • FREEDOM TORIC EdoF, EdOf Toric with Clear Hydrophobic
  • FREEDOM TORYC EdOf, EdoF Toric with Yellow Hydrophobic
  • FREEDOM TRITORIC, Trifocal Toric with Clear Hydrophobic
  • FREEDOM TRITORYC, Trifocal Toric with Yellow Hydrophobic

Our MULTIFOCAL TORIC Hydrophobic Lenses provides both NEAR and FAR vision for ASTIGMATIC CATARACT PATIENTS who desire spectacle independency. It combines the CYLINDER technology of a TORIC lens in the anterior side and the Diffractive-Refractive Element of a BIFOCAL lens in the posterior side.

Our EDOF TORIC Hydrophobic Lenses provides both NEAR, INTERMEDIATE and FAR vision for ASTIGMATIC CATARACT PATIENTS who desire spectacle independency. It combines the CYLINDER technology of a TORIC lens in the anterior side and the Diffractive-Refractive Element of a EdOf lens in the posterior side.

Our TRIFOCAL TORIC Hydrophobic Lenses provides both NEAR, INTERMEDIATE and FAR vision for ASTIGMATIC CATARACT PATIENTS who desire spectacle independency. It combines the CYLINDER technology of a TORIC lens in the anterior side and the Diffractive-Refractive Element of a TRIFOCAL lens in the posterior side.

Freedom provides large DIOPTER range of astigmatism correction IOLs with custom-made PRECISION PERFECTION. Spherical power from 10.0D to 30.0D and cylinder power range is up to 12.0D with half diopter increments (monofocal lens only with a bi-toric premium technology).

Following cylinder options are available to suit the requirement of any patients :

IOL Cylinder Power Recommended Corneal Astigmatism Correction ranges
1.50D 0.75 – 1.5D
2.25D 1.50-2.00D
3.00D 2.00-2.50D
3.75D 2.50-3.00D
4.50D 3.00-3.50D
5.25D 3.50D-4.00D
6.00D 4.00D-4.50D
6.75D (monofocal only with bi-toric optic) 4.50D-5.00D
7.50D (monofocal only with bi-toric optic) 5.00D-5.50D
8.25D (monofocal only with bi-toric optic) 5.50D-6.00D
9.00D (monofocal only with bi-toric optic) 6.00D-6.50D
9.75D (monofocal only with bi-toric optic) 6.50D-7.00D
10.50D (monofocal only with bi-toric optic) 7.00D-7.50D
11.25D (monofocal only with bi-toric optic) 7.50D-8.00D
12D (monofocal only with bi-toric optic) 8.50D and above

Toric Sucess Parameters

Patient with regular pre-operative astigmatism with an intact capsular bag, and no pre-existing ocular pathology who desire spectacle independency are good candidates for Toric surgery. Patient with keratoconus, or irregular astigmatism are recommended against Toric lens.

Toric IOL Power calculation: Connect to www.freedomtoriccalculator.com
⦁ SPHERICAL POWER – Is determined by using the preferred method such as optical biometry as for conventiol IOLs using personalized A-constant.
⦁ CYLINDRICAL POWER- Steep K & Flat K values are determined through Keratometry and corneal topography for corneal measurements
⦁ With FREEDOM TORIC CALCULATOR, determine the appropriate Toric IOL model, power, & axis of placement using default SIA measurements: 0.5 D if your incision is 2.5 mm or smaller.

Preoperative Marking: Reference marks are made at 3, 6 and 9’o clock in upright position to limit cyclotorsional effect using marking tools such as level/bubble marker or free-hand marker.

Intraoperative Marking: Calculated steep axis is marked using fixation rings as Mendez at two positions (180° apart) with the guide of reference axis in supine position for alignment of Toric IOL.

Primary/GROSS IOL Alignment: After a standard cataract implantation in the capsular bag, manipulate the IOL into place within 15 to 20 degrees short of the intended axis. Remove all the residual viscoelastics gently with care.

Final IOL alignment: Rotate clockwise to align with intended axis on the cornea.

Freedom TORIC
Freedom TORYC

Hydrophobic Acrylic Posterior Chamber Toric Aspheric 360° Square Edge Single Piece Foldable Intraocular Lens

Model TPC602SQ TPC602SQP
Design
Optic Design Equiconvex - 360° Square Edge with Toric Aspheric Optic
Optic Size 6.00mm
Over All length 13.00mm
Estimated A-Constant 118.40
Recommended Optical A-Constant A Constant SRK/T : 119.20 | A Constant SRK/II : 119.50
Hoffer Q : 5.65 | Holladay 1 : 1.87 | Holladay 2 : 5.199 | Haigis : a0=1.441/a1=0.400/a2=0.100
Diopter Range 5.00 D to 30.00 D (0.5 Diopter increments)
AC Depth 5.00mm
Refractive Index 1.493
Cartridge Size FDC220_60DEG

Yellow Hydrophobic Acrylic Posterior Chamber Toric Aspheric 360° Square Edge Single Piece Foldable Intraocular Lens

Model TPC602SQY TPC602SQYP
Design
Optic Design Equiconvex - 360° Square Edge with Toric Aspheric Optic
Optic Size 6.00mm
Over All length 13.00mm
Estimated A-Constant 118.40
Recommended Optical A-Constant A Constant SRK/T : 119.20 | A Constant SRK/II : 119.50
Hoffer Q : 5.65 | Holladay 1 : 1.87 | Holladay 2 : 5.199 | Haigis : a0=1.441/a1=0.400/a2=0.100
Diopter Range 5.00 D to 30.00 D (0.5 Diopter increments)
AC Depth 5.00mm
Refractive Index 1.493
Cartridge Size FDC220_60DEG